Ultrasonic Flaw Detector
-
Ultrasonic Thickness Gauge
-
Coating Thickness Gauge
-
Portable Hardness Tester
-
X-Ray Flaw Detector
-
X-ray Pipeline Crawlers
-
Holiday Detector
-
Magnetic Particle Testing
-
Eddy Current Testing Equipment
-
Penetrant Testing
-
Shore Durometer
-
Brinell Hardness Tester
-
Rockwell Hardness Tester
-
Micro Vickers Hardness Tester
-
Material Hardness Tester
-
Vibration Meter
-
Surface Roughness Tester
-
Profile Projector
-
Gloss Meters
-
Concrete Test Hammer
-
Laser Infrared Thermometer
-
bobAstm E114-15 Inch Type 1018 Calibration Test Block Carbon Steel Mini Iiw 2
-
WilliamAstm E114-15 Inch Type 1018 Calibration Test Block Carbon Steel Mini Iiw 2
-
tomAstm E114-15 Inch Type 1018 Calibration Test Block Carbon Steel Mini Iiw 2
-
LeoAstm E114-15 Inch Type 1018 Calibration Test Block Carbon Steel Mini Iiw 2
Digital ASTMD2240 / JISK7215 0 - 100HC Shore Durometer HT-6510C

Contact me for free samples and coupons.
WhatsApp:0086 18588475571
Wechat: 0086 18588475571
skype: sales10@aixton.com
If you have any concern, we provide 24-hour online help.
xMeasurement Range | 0-100HC | Measurement Deviation | <1%H |
---|---|---|---|
Power Supply | 4x1.5V AAA (UM-4) Battery | Battery Indicator | Low Battery Indicator |
Highlight | digital shore durometer,portable shore durometer |
Weight (not including probe):173g
Standard accessories:
Main unit
Zero Calibration plate
Carrying case
Instruction Manual
Certificate
CALIBRATION CHECK
To check whether the tester is accurate, just Insert the indenter into the hole of the calibrated test Block. Apply enough force to make firm contact between the top surface of the test block and the presser foot. The reading
should agree with the value stamped on the test block. If not, just carry out Zero calibration and High end calibration.
Zero calibration
Hold the durometer vertically with the point of the indenter hanging in the air, the reading on the display should be '0'. If not, depress the 'ZERO' key to make the tester display '0' .
High end calibration
Just place the indenter onto a flat glass, apply enough force to make firm contact between the glass and the presser foot. The readings on the display should lie between 99.5 and 101. If not, press CAL key to carry out high end calibration.
BATTERY REPLACEMENT
When the battery symbol appears on the display, it is time to replace the batteries.
Slide the Battery Cover away from the tester and remove the batteries.
Install batteries paying careful attention to polarity.
NOTE
Readings below 10HD for Shore D type may be inexact and should not be reported for some materials. Measurements should be made on a Shore A type.
Readings above 90HA for Shore A should be made on a Shore D type durometer